PROBE:
Standard 316L Stainless Steel or Hastelloy probe straw
EPA approved lengths
2 micron filter standard, other sizes available
Heated probe filter maintained above 250°F [121°C]
Insulating jacket provided
PROBE SUPPORT BUNDLE:
¼" [.95 cm] Teflon calibration line
½" [0.64 cm] Polyethylene backpurge line
14 AWG triad for probe heater power
Fire retardant jacket
HEATED SAMPLE LINE:
3/8" [0.95 cm] Teflon or 316 Stainless Steel sample tube
Recommended maximum length 250 feet [76.2 meter]
Line temperature controlled to 300°F [148°C]
Low temperature alarm provided to Controller
Insulated fire retardant jacket
SAMPLE CONDITIONER:
Thermo-electric cooler
Negligible loss of NO2 and SO2
Teflon diaphragm pump
Peristaltic moisture drain pump
2 micron Teflon sample filter
Water slip alarm, sample dew point alarm
Probe and sample line vacuum alarm
CABINET VERSION:
Single bay, 19" [48.3 cm] Rack Mount
Standard NEMA 12 with fan ventilation or HVAC.
Other NEMA ratings are available.
Dimensions 81"H x 25"W x 39"D [2.0 x 0.6 x 1.0 meter] Approx. 400lbs.[181Kg]
Front door and rear doors, removable side panels, sight glass available
Lifting lugs and/or castors can be provided
OPEN ARCHITECTURE VERSION:
Open rack with optional desk.
Dimensions 80"H x 63"W x 30"D [2.0 x 1.6 x 0.8 meter] Approx. 300lbs. [136Kg]
Sample Conditioning Plate.
28"H x 24"W x 9"D [0.7 x 0.6 x 0.2 meter] Approx. 60 lbs. [27Kg]
SYSTEM CONTROLLER:
GE or Allen Bradley Controller, C3io provides data buffering, digital inputs, digital outputs, analog inputs, analog outputs, and Ethernet communications. Optional HMI provided for manual system control.
GAS ANALYZERS:
Model Number T300, for CO
Model Number T200, for NOX
Model Number T100, or SO2
Model Number T201, for NH3
Model Number T360, T801, T802, T803 for O2/CO2
RANGES:
CO 0-5000ppm
NOx 0-5000ppm
SO2 0-5000ppm
O2 0-5/10/25%
CO2 0-2000 to 0-3000, 0-10/15/20%
NH3 0-20ppm
RELATIVE ACCURACY:
± 20% of Reference Methods for 40 CFR 60
± 7.5% (better than ± 10.0%) of Reference Methods for 40 CFR 75
RESPONSE TIME:
Less than 90 seconds to 95% full scale from system inlet
CALIBRATION DRIFT:
Zero and Span
± 0.5% absolute for O2/CO2%
± 2.5% of full scale for CO, NOX, NH3, and SO2 ppm
REPEATABLILITY: ± 1% of full scale
PROCESS: 80-1800°F [27-982°C] , up to 50% moisture
AMBIENT: Varies by Engineered Enclosure
SYSTEM UTILITIES: -40°F [-40°C] dew point instrument air, 80 psig [551.6 KPa] minimum, 0.5 SCFM [14 lpm] , Single point attachment of 240Vac, single phase, 60 hertz, 50 amp power
VALUE ADDED OPTIONS:
- Regulatory Consultation
- LightHawk® 560 Opacity Monitor integration
- LaserHawk® 360 Particulate Monitor integration
- Ultraflow 150 Ultrasonic Stack Flow integration
- Automatic Linearity's
- RegPerfect® Data Acquisition System
- Training
- Installation supervision
- Startup
- EPA Certification
- Maintenance Contracts